Biochemical Analyzer Filters
Narrow bandpass filters for photometric and turbidimetric channels
PB01 · IMAGE PENDINGColored narrow bandpass filters for analyzersNarrow Bandpass Filters for Clinical Chemistry Photometers
A biochemical analyzer measures how much light a sample absorbs at a few fixed wavelengths. Each wavelength needs a narrow bandpass filter in front of the detector.
Our analyzer filters cover 12 wavelengths from 340 to 670 nm. The 340 nm filter is 10 nm wide, the others 8 nm, all ±2 nm.
Standard filters transmit at least 85% or 90% and block at OD4, OD5 or OD6, chosen to suit your instrument.
Energy-matched sets go further: each channel’s transmission is set to balance your lamp and detector across the wavelengths.
Other centers from 340 to 800 nm and sizes to your filter wheel are made to order. Specify a custom filter
Standard or Energy-Matched?
D01 · IMAGE PENDINGStandard versus energy-matched channel transmissionStandard Filters
Every filter has the same specification: transmission of at least 85% or 90%, 8 or 10 nm wide, OD4 to OD6. Choose these when your instrument already corrects for differences between channels in software.
Energy-Matched Sets
Each channel’s transmission is set to match the output of your lamp and the response of your detector at that wavelength, so all channels give similar signal levels. Typical wavelengths: 340, 405, 450, 510, 546, 578, 630 and 670 nm.
Why Blocking Matters in Absorbance Measurements
Any light outside the band adds stray signal and makes strong absorbances read too low. The deeper the blocking, the higher the absorbance your analyzer can measure accurately.
The 340 nm channel is the hardest. Halogen lamps and silicon detectors give little signal in the UV and much more in the visible, so a 340 nm filter must block the visible deeply. Choose OD5 or OD6 there.
Blocking extends from 200 to 1,100 nm on our standard analyzer filters.
How We Match a Set to Your Instrument
Send us your lamp’s spectrum, your detector’s responsivity and the wavelengths you use. We set the transmission of each filter so the signal reaching the detector is balanced across the channels.
The strong channels get lower transmission and the weak ones, usually in the UV, keep the highest. The result is a set that uses your detector’s range evenly.
Each energy-matched set is supplied with the measured transmission of every channel.
Standard Analyzer Filters
| Filter | Center (nm) | FWHM (nm) |
|---|---|---|
| NBP340 | 340 ± 2 | 10 ± 2 |
| NBP405 | 405 ± 2 | 8 ± 2 |
| NBP450 | 450 ± 2 | 8 ± 2 |
| NBP492 | 492 ± 2 | 8 ± 2 |
| NBP505 | 505 ± 2 | 8 ± 2 |
| NBP510 | 510 ± 2 | 8 ± 2 |
| NBP546 | 546 ± 2 | 8 ± 2 |
| NBP578 | 578 ± 2 | 8 ± 2 |
| NBP600 | 600 ± 2 | 8 ± 2 |
| NBP630 | 630 ± 2 | 8 ± 2 |
| NBP660 | 660 ± 2 | 8 ± 2 |
| NBP670 | 670 ± 2 | 8 ± 2 |
Transmission at least 85% or 90%; blocking OD4, OD5 or OD6, specified per order. Ask us for the curves.
Designing or servicing an analyzer? Send us your wavelengths, lamp, detector and filter size, and we will reply with standard filters or an energy-matched set and a quote. Contact us