Flow Cytometry Filters

Bandpass and dichroic filters for every laser and detector channel

Mounted Bandpass Filters for Flow Cytometry

Bandpass Filters for Flow Cytometer Scatter and Fluorescence Channels

A flow cytometer measures light scattered and emitted by cells passing one or more lasers. Each detector needs a bandpass filter that passes only its signal.

Our flow filters transmit at least 90% and block at OD6, which keeps scattered laser light out of the fluorescence channels.

Standard filters cover the 488 and 638 nm laser lines, 10 nm wide, and emission bands at 460, 535, 580 and 665 nm.

Widths from 10 to 80 nm and other centers are made to order, with matching dichroic mirrors.

Flow filters reach up to 98% transmission at the design wavelength, with edges under 1% slope, so neighbouring channels stay apart.

Mounted Flow Cytometry FiltersP01 · IMAGE PENDINGMounted flow cytometry filters

Sizes are cut to your detector module. Specify a custom filter

Filters by Laser Line

Blue laser, 488 nm. A 488/10 filter passes 488 ± 5 nm, the usual choice for side-scatter detection or for cleaning the laser line. 535/40 and 580/20 filters sit in the common green and yellow detection windows, near FITC and PE.

Red laser, 638 nm. A 638/10 filter passes the red laser line; a 665/20 filter, 655 to 675 nm, sits in the common window for APC-type dyes.

Flow Cytometer PathsD01 · IMAGE PENDINGFlow cytometer light path
Figure: Flow cytometer light path.

In most cytometers, a chain of dichroic mirrors splits the emission by wavelength, and each detector then has its own bandpass filter. Dichroic mirrors

If your instrument uses other lasers, such as 405, 561 or 640 nm, send us the line and detection bands. We design to order.

Violet laser. A 460/55 filter, about 433 to 488 nm, covers the common blue detection window for violet-excited dyes.

Reference transmission curves for a family of visible bandpass filtersT01 · IMAGE PENDINGMeasured curves of four flow cytometry filters

The visible bandpass curves illustrate how separate wavelength channels can be selected across the spectrum.

Figure: Measured curves of four flow cytometry filters.

Why Flow Filters Need Deep Blocking

Laser light scattered by cells is far stronger than their fluorescence. A filter for the 535 nm channel must pass 515 to 555 nm and block the 488 nm laser just 27 nm below its band.

Our flow filters block at OD6, which lets through one part in a million of the light outside the band. Blocking extends from 200 to 1,100 nm.

Check that each filter also blocks the other lasers in your instrument, not only its own. A filter in a 488 nm channel can also see 638 nm light.

Standard Flow Cytometry Filters

Filter Center (nm) FWHM (nm) Approx. passband (nm) Common use
BP460/55 460 ± 2 55 ± 2 433–488 Violet-laser emission
BP488/10 488 ± 2 10 ± 2 483–493 488 nm line / side scatter
BP535/40 535 ± 2 40 ± 2 515–555 488 nm-excited emission
BP580/20 580 ± 2 20 ± 2 570–590 488 nm-excited emission
BP638/10 638 ± 2 10 ± 2 633–643 638 nm line
BP665/20 665 ± 2 20 ± 2 655–675 638 nm-excited emission

All transmit at least 90% with OD6 blocking. Passbands are estimated from center and FWHM.

What to Send Us

  • Laser lines in your instrument
  • Center and width for each detection channel
  • Blocking OD and range, including other lasers
  • Filter size, thickness and mounting
  • Dichroic mirrors needed in the splitting chain
  • Annual quantity

Building or servicing a flow cytometer? Send us its laser lines and detection channels, and we will reply with matching bandpass filters, dichroics and a quote. Contact us