Dichroic 45D-HT340/HR525-940

45° UV dichroic: over 90% transmission at 340 nm, under 5% from 525 to 940 nm.

Need a different size or spec? We make custom versions.

BANDOPTIC spectrum engine

Interactive Spectrum

Plot and compare optical filter spectra. Catalog curves digitized from the supplied PDFs are approximate references; request measured data for specifications.

Choose a filter from the catalog or upload a numerical CSV.

Request measured data

Catalog curves are approximate digital tracings of PDF graph images, not instrument CSV or guaranteed production-lot results. Some curves show only the plotted peak or cut-on edge; open the source graph for its full context. For approximate traces, OD values at or below 2% transmission are hidden because raster plots cannot verify blocking performance. Uploaded CSV files stay in your browser. Andover numerical curves will be added only when source data is available.

Technical Specifications

SpecificationValue
Transmission> 90% minimum at 340 nm
Reflect bandT < 5% (average), 525–940 nm
50% edgeBetween 340 and 525 nm; see the measured curve
Angle of incidence45°
Cone half-angle3°
SubstrateFused silica (JGS1)
Thickness1.0 ± 0.1 mm
Dimensional tolerance+0 / −0.1 mm
Surface quality60-40 scratch-dig
Clear aperture≥ 90% of diameter

Sizes

Need custom requirements? We cut to your drawing.

SizeThicknessMount
Ø12.5 mm1.0 mmUnmounted
Ø25 mm1.0 mmUnmounted or ringed
12.5 × 12.5 mm1.0 mmUnmounted

Other sizes and shapes are cut to order.

Choosing This Filter

This 45° dichroic separates ultraviolet from visible and near-infrared light. It transmits at least 90% at 340 nm and keeps transmission under 5% on average from 525 to 940 nm, so a UV channel goes straight through while green, red and near-infrared light is folded out of the path. That suits instruments that combine a UV measurement at 340 nm, common in photometric analyzers, with visible channels in the same optical train. The 50% point between 340 and 525 nm is not yet specified, so ask us for the measured curve. For a dichroic that reflects UV and passes visible, ask about our HR300-380/HT420-700 design.

Customization Options

Need a different wavelength, width or blocking? We design it to order.

We can move the center wavelength anywhere from 200 to 2,500 nm, change the width, block deeper, up to OD6, or further into the infrared, and cut the filter to your size and thickness.

Tell us your light source, your detector and the numbers you need. We will reply with a recommended design and a quote, from a single prototype to production volumes.

See what we can make on our custom filters page. Learn more

FAQs

Can I tilt this filter?

It is designed for 45°, the usual angle when a beam is split or combined. Its edges move with angle and are different for s- and p-polarized light, so mount it at 45° and tell us if your design uses another angle or a converging beam.

How should I clean and store it?

Hold the filter by its edges. Blow off dust first, then wipe gently with lens tissue and reagent-grade isopropyl alcohol if needed. Store it dry, in its case, away from direct sunlight.

Can I get the measured curve with my order?

Yes. Ask for it when you order, and we include the measured transmission of your batch, as a chart and as a data file.

Don’t see the filter you need? Tell us about your system and we will suggest one or design it. Learn more