DM770

Longpass dichroic for filter sets: at least 95% reflection from 600 to 755 nm, 95% transmission from 780 to 900 nm at 45°.

Need a different size or spec? We make custom versions.

BANDOPTIC spectrum engine

Interactive Spectrum

Plot and compare optical filter spectra. Catalog curves digitized from the supplied PDFs are approximate references; request measured data for specifications.

Choose a filter from the catalog or upload a numerical CSV.

Request measured data

Catalog curves are approximate digital tracings of PDF graph images, not instrument CSV or guaranteed production-lot results. Some curves show only the plotted peak or cut-on edge; open the source graph for its full context. For approximate traces, OD values at or below 2% transmission are hidden because raster plots cannot verify blocking performance. Uploaded CSV files stay in your browser. Andover numerical curves will be added only when source data is available.

Technical Specifications

SpecificationValue
Transmission≥ 95% (average), 780–900 nm
Reflection≥ 95%, 600–755 nm
Angle of incidence45°
Cone half-angle3.5°
SubstrateOptical glass (K9 / B270)
Thickness1.0 ± 0.1 mm
Dimensional tolerance+0 / −0.1 mm
Surface quality60-40 scratch-dig
Clear aperture≥ 90% of diameter

Sizes

Need custom requirements? We cut to your drawing.

SizeThicknessMount
Ø12.5 mm1.0 mmUnmounted
Ø25 mm1.0 mmUnmounted or ringed
12.5 × 12.5 mm1.0 mmUnmounted

Other sizes and shapes are cut to order.

Choosing This Filter

DM770 is the dichroic mirror in our Cy7 PCR set, the longest-wavelength channel we offer. At 45° it reflects at least 95% from 600 to 755 nm, covering excitation light around 725 nm, and transmits at least 95% on average from 780 to 900 nm for near-infrared emission around 800 nm. Its reflect band starts at 600 nm rather than 450 nm, which is enough for this channel’s excitation. Sold on its own for PCR and near-infrared fluorescence optics. For the Cy5.5 channel, see DM700.

Customization Options

Need a different wavelength, width or blocking? We design it to order.

We can move the center wavelength anywhere from 200 to 2,500 nm, change the width, block deeper, up to OD6, or further into the infrared, and cut the filter to your size and thickness.

Tell us your light source, your detector and the numbers you need. We will reply with a recommended design and a quote, from a single prototype to production volumes.

See what we can make on our custom filters page. Learn more

FAQs

Can I tilt this filter?

It is designed for 45°, the usual angle when a beam is split or combined. Its edges move with angle and are different for s- and p-polarized light, so mount it at 45° and tell us if your design uses another angle or a converging beam.

How should I clean and store it?

Hold the filter by its edges. Blow off dust first, then wipe gently with lens tissue and reagent-grade isopropyl alcohol if needed. Store it dry, in its case, away from direct sunlight.

Can I get the measured curve with my order?

Yes. Ask for it when you order, and we include the measured transmission of your batch, as a chart and as a data file.

Don’t see the filter you need? Tell us about your system and we will suggest one or design it. Learn more