LP1660
Longpass filter with a 1,660 nm cut-on: OD3 blocking from 200 to 1,625 nm, and more than 92% transmission from 1,700 to 2,500 nm.
Need a different size or spec? We make custom versions.
BANDOPTIC spectrum engine
Interactive Spectrum
Plot and compare optical filter spectra. Catalog curves digitized from the supplied PDFs are approximate references; request measured data for specifications.
Catalog curves are approximate digital tracings of PDF graph images, not instrument CSV or guaranteed production-lot results. Some curves show only the plotted peak or cut-on edge; open the source graph for its full context. For approximate traces, OD values at or below 2% transmission are hidden because raster plots cannot verify blocking performance. Uploaded CSV files stay in your browser. Andover numerical curves will be added only when source data is available.
Technical Specifications
| Specification | Value |
|---|---|
| Cut-on wavelength | 1,660 ± 2 nm |
| Transmission | > 92% (average), 1,700–2,500 nm |
| Blocking | OD > 3 (average), 200–1,625 nm |
| Angle of incidence | 0 ± 2° |
| Cone half-angle | 3.5° |
| Substrate | Optical glass (K9 / B270) |
| Thickness | 1.0 ± 0.1 mm |
| Dimensional tolerance | +0 / −0.1 mm |
| Surface quality | 60-40 scratch-dig |
| Clear aperture | ≥ 90% of diameter |
Sizes
Need custom requirements? We cut to your drawing.
| Size | Thickness | Mount |
|---|---|---|
| Ø12.5 mm | 1.0 mm | Unmounted |
| Ø25 mm | 1.0 mm | Unmounted or ringed |
| 12.5 × 12.5 mm | 1.0 mm | Unmounted |
Other sizes and shapes are cut to order.
Choosing This Filter
LP1660 blocks to OD3 from 200 to 1,625 nm and passes more than 92% on average from 1,700 to 2,500 nm, with its cut-on at 1,660 ± 2 nm. That wide pass band covers the long end of the short-wave infrared, beyond what standard InGaAs detectors see, so it pairs with extended InGaAs, lead-salt and other detectors that reach toward 2.5 µm. It keeps visible light, 1,064 nm and 1,550 nm laser lines and the rest of the shorter SWIR out of the channel. Choose it for SWIR spectroscopy and sensing above 1.7 µm. For a cut-on closer to 1.8 µm with deeper focus on 2 µm, see LPF1755.
Customization Options
Need a different wavelength, width or blocking? We design it to order.
We can move the center wavelength anywhere from 200 to 2,500 nm, change the width, block deeper, up to OD6, or further into the infrared, and cut the filter to your size and thickness.
Tell us your light source, your detector and the numbers you need. We will reply with a recommended design and a quote, from a single prototype to production volumes.
See what we can make on our custom filters page. Learn more
Related Filters
FAQs
Can I tilt this filter?
A small tilt, up to about 2°, stays within spec. Beyond that, the edge moves to shorter wavelengths, and the further you tilt, the further it moves. If your design needs a tilt, tell us the angle and we will design for it.
How should I clean and store it?
Hold the filter by its edges. Blow off dust first, then wipe gently with lens tissue and reagent-grade isopropyl alcohol if needed. Store it dry, in its case, away from direct sunlight.
Can I get the measured curve with my order?
Yes. Ask for it when you order, and we include the measured transmission of your batch, as a chart and as a data file.
Don’t see the filter you need? Tell us about your system and we will suggest one or design it. Learn more