Multiphoton SP520

520 nm shortpass emission filter for multiphoton microscopy: more than 90% transmission from 400 to 510 nm, and OD2 blocking from 530 to 1,100 nm.

Need a different size or spec? We make custom versions.

BANDOPTIC spectrum engine

Interactive Spectrum

Plot and compare optical filter spectra. Catalog curves digitized from the supplied PDFs are approximate references; request measured data for specifications.

Choose a filter from the catalog or upload a numerical CSV.

Request measured data

Catalog curves are approximate digital tracings of PDF graph images, not instrument CSV or guaranteed production-lot results. Some curves show only the plotted peak or cut-on edge; open the source graph for its full context. For approximate traces, OD values at or below 2% transmission are hidden because raster plots cannot verify blocking performance. Uploaded CSV files stay in your browser. Andover numerical curves will be added only when source data is available.

Technical Specifications

SpecificationValue
Cut-off wavelength520 ± 2 nm
Transmission> 90% (average), 400–510 nm
BlockingOD > 2 (average), 530–1,100 nm (design value)
Angle of incidence0°
Cone half-angle3.5°
SubstrateOptical glass (K9 / B270)
Thickness1.0 ± 0.1 mm
Dimensional tolerance+0 / −0.1 mm
Surface quality60-40 scratch-dig
Clear aperture≥ 90% of diameter

Sizes

Need custom requirements? We cut to your drawing.

SizeThicknessMount
Ø12.5 mm1.0 mmUnmounted
Ø25 mm1.0 mmUnmounted or ringed
12.5 × 12.5 mm1.0 mmUnmounted

Other sizes and shapes are cut to order.

Choosing This Filter

Multiphoton SP520 passes more than 90% on average from 400 to 510 nm and blocks at OD2 from 530 to 1,100 nm, with its edge at 520 ± 2 nm. In a multiphoton microscope it collects blue and blue-green emission and rejects green, red and near-infrared light. OD2 is enough to split channels but not to hold back a femtosecond laser, so pair it with a dedicated laser-blocking filter such as Multiphoton SP800. The OD values beyond the edge are design values. For a channel that stops at 480 nm with deeper blocking just above it, see Multiphoton SP490.

Customization Options

Need a different wavelength, width or blocking? We design it to order.

We can move the center wavelength anywhere from 200 to 2,500 nm, change the width, block deeper, up to OD6, or further into the infrared, and cut the filter to your size and thickness.

Tell us your light source, your detector and the numbers you need. We will reply with a recommended design and a quote, from a single prototype to production volumes.

See what we can make on our custom filters page. Learn more

FAQs

Can I tilt this filter?

A small tilt, up to about 2°, stays within spec. Beyond that, the edge moves to shorter wavelengths, and the further you tilt, the further it moves. If your design needs a tilt, tell us the angle and we will design for it.

How should I clean and store it?

Hold the filter by its edges. Blow off dust first, then wipe gently with lens tissue and reagent-grade isopropyl alcohol if needed. Store it dry, in its case, away from direct sunlight.

Can I get the measured curve with my order?

Yes. Ask for it when you order, and we include the measured transmission of your batch, as a chart and as a data file.

Don’t see the filter you need? Tell us about your system and we will suggest one or design it. Learn more